1-52-removebg-preview.png
+
  • 1-52-removebg-preview.png

Pressure flow test

Product model:


Classification:

Testing Equipment

Pressure flow test equipment is used to test the pressure resistance of catheters, balloons, conveyors and other products under specific requirements. The pressure module is mainly used to test the physical properties of the product such as pressure resistance, fatigue and compliance under set pressure conditions. The flow module is mainly used to test the physical properties of the product when injected into a set liquid volume. The operator can set the program and complete the operation through the touch screen, or set the program on the computer, the test curve is displayed on the computer in real time, and the test results can be saved and exported.

Key words: Balloon line

Contact Us Messages

Product Details


Pressure flow testing equipment is used to test the pressure resistance of catheters, balloons, conveyors and other products under specific requirements. The pressure module is mainly used to test the physical properties of the product such as pressure resistance, fatigue, and compliance under set pressure conditions. The flow module is mainly used to test the physical properties of the product when injected into a set volume of liquid. The operator can set the program and complete the operation through the touch screen, or set the program on the computer, the test curve is displayed on the computer in real time, and the test results can be saved and exported as a document.

 

Equipment features:

◆ Fatigue test

◆ Withstand voltage test

◆ Compliance test

◆ Volume pressure test

◆ Volumetric anterograde test

◆ Volumetric fatigue test

◆ High test accuracy

◆ Real-time display of curves

◆ Parameters can be stored

◆ Easy to operate

 

Technical Specifications:

 

Pressure module

Volumetric flow module

other

Pressure range

0-1000 psi

Volumetric flow rate

0.1-20ml/s

power supply

220V,50HZ

Pressure accuracy

Accuracy: ±3 psi

Pressurization/depressurization rate

0.1-20ml/s

power

1000W

resolution

1 psi

Leakage rate

0.3 atm/s

Compressed air input

0.4-0.7Mpa

Temperature range

Room temperature -50 °C

Volume mode pressure range

0-200psi

fountainhead

Purified water or water for injection

Temperature accuracy

±3°C

Volumetric mode pressure accuracy

±1 psi

Use ambient humidity

30-70%RH

Temperature resolution

0.1℃

Volume of water for injection

200ml

Device dimensions

700X70x400mm

Diameter range

0-0.5"

Volumetric accuracy

±0.5ml

weight

50 Kg

Diameter accuracy

0.001"

Volumetric resolution

0.1ml

 

 

Diameter resolution

0.001"

 

 

 

 

 